ABS Compact Built-in Certified Vaccine Refrigerators

Category Benchtop / Undercounter Refrigerators
Manufacturer American BioTech Supply
Country of manufacture the United States of America
Overview

These 2.5 Cubic Feet premier built-in undercounter refrigerators are designed in accordance with the NSF/ANSI 456* Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

These refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

*-one or more of these certifications may apply to this unit.

NOTE: American Biotech Supply refrigerators and freezers are manufactured at the time the order is placed then go through extensive testing to ensure quality. A lead time of approximately 3 weeks should be expected.

Key Attributes 
Min. Temperature 1°C to 10°C
Cubic Feet 2.5
Alarms Yes
Controls Digital
Defrost Cycle optimized, zero energy
Electrical 110-120v ac: 15 a
Warranty (years) 2
Weight (lb) 88
Width (in) 17 3/4" 
Depth (in) 20 7/8" 
Height (in) 30 3/4" 
Manufacturer Specifications

Description    Single Solid Door Pharmacy/Vaccine Undercounter Refrigerator Built-In
Operational Environment    Indoor Use Only, +18°C To +26°C (+65°F To +78°F), <70% Rh
Storage Capacity    2.5 Cu. Ft. Gross Volume
Door    One Swing Solid Door, Self-Closing, Right Hinged, Non-Reversible, Magnetic Sealed Gasket, Keyed Lock
Shelves    Four Shelves (Three Adjustable/One Fixed) With Guard Rail On Back
Mounting    Leveling Legs
Interior Lighting    N/A
Airflow Management    Forced Air Technology, Patent Pending
External Probe Access    Side Wall Port (3/8″) Dia.
Insulation    Cabinet Is Foamed-In-Place With Epa Compliant High Density Urethane Foam
Exterior Materials    White Powder Coated Steel
Access Control    Pyxis®, Omnicell® And Acudose Rx® Compatible
General Warranty    Two (2) Years Parts And Labor Warranty, Excluding Display Probe Calibration
Compressor Warranty    Five (5) Years Compressor Warranty
Product Weight    72 Lbs.
Shipping Weight    88 Lbs.
Rated Amperage    0.9 Amps
Power Plug/Power Cord    Nema 5-15 Plug, 8 To 10 Ft Typical, Conforms To Ul471 Requirements, Vaccine Storage Power Cord Warning Label
Facility Electrical Requirement    110-120V Ac: 15 A (Minimum)
Agency Listing And Certification    Certified In Accordance With The Nsf/Ansi 456 Standard For Vaccine Storage. Ul, C-Ul, Etl, C-Etl Listed (Either Single Or Dual Agency Listings) And Certified To Ul471 Standard, Hydrocarbon Refrigerant Safety.
Included Accessories    Temperature Monitor Device (Tmd) Complies With The Current Cdc Guidelines, With 3 Years Certification Of Calibration, “Buffered” Probe In The Product Simulated Solution, Min/Max Memory, Field Installable, And Visual & Audible Temp Alarm
Uniformity¹ (Cabinet Air)    +/-1.1°C
Stability² (Cabinet Air)    +/- 0.7°C
Maximum Temperature Variation (Cabinet Air)    +/- 1.2°C
Temperature Rise After 8 Sec Door Openings    Temperature Did Not Exceed 7.2°C At Any Probe For All Required Nsf/Ansi 456 Testing Protocolsᶟ
Recovery After 3 Min Door Opening    All Probes Recover To Under 8°C Within 7.5 Min.
Energy Consumption    0.58 Kwh/Day⁴
Average Heat Rejection    1.00 Kwh/Day (142 Btu/H)⁴
Noise Pressure Level (Dba)    34 Or Less Installed
Pull Down Time To Nominal Operating Temp    42 Min
Performance Data Acquired At 22°C Ambient, Using Nsf/Ansi 456 Compliant Validation Ballast Probes, Empty Chamber, During Stabilized Steady State Operation And A Daq Sampling Rate Of One Measurement Every 10 Seconds    
1 – Uniformity Is Defined As The Maximum Variance In Temperature Across All Probes At Any Point In Time Over The Testing Period    
2 – Stability Is Defined As The Maximum Variance In Temperature Experienced By Any Single Probe Over The Testing Period    
3 – Temperature Performance For All Loaded And Unloaded Door Opening Protocols, All Alarm, Controller And Probe Requirements As Defined In The Nsf/Ansi 456 Standard For Vaccine Storage    
4 – Data Per Energy Star Test Results Or Equivalent Testing And Calculation. Heat Rejection Based On Daily Averages, Not Continuous Operation. Performance Exceeds Energy Star Requirements.    
Controller Technology    Parametric, Microprocessor, Led Display With 0.1°C Resolution
Temperature Setpoint Range    1°C To 10°C (Setpoint Must Remain Unaltered From The Factory Setting To Remain Compliant With Nsf/Ansi 456 Standard For Vaccine Storage Requirements)
Display Probe    Calibrated, Stainless Steel
External Alarm Connection    State Switching Remote Alarm Contacts
Alarms    Visual And Audible Indicators
Alarms    High / Low Temperature, Compliant With Alarm Requirements Defined In The Nsf/Ansi 456 Standard For Vaccine Storage
Simulator Ballast    Glass Bead Thermal Media